inforMED
MalfunctionFZP

LIGATURE CLIP 12 MAG.=144 PCS.

Received Aug 24, 2015 · Event occurred Jan 27, 2015

Report 2916714-2015-00758 · MDR key 5044874

Device

Manufacturer

Aesculap, Inc.

Model number

PL572T

Catalog number

PL572T

Lot number

52064447

Product problems

  • Defective Device

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

US REPORTING AGENT NOTIFIED ON: (B)(6)2015. MANUFACTURING SITE EVALUATION: MOUNTING CLIP AT THE BACK IS BROKEN OFF; THE BROKEN FRAGMENT WAS NOT RECEIVED FOR EVALUATION. VISUAL MICROSCOPIC EVALUATION SHOWS THE BREAKAGE SURFACE IS A FORCED FRACTURE DUE TO OVERLOAD. THERE WERE NO PORES, INCLUSIONS, OR FOREIGN BODIES FOUND AT THE POINT OF RUPTURE. MANUFACTURING RECORDS WERE REVIEWED FOR THE REPORTED LOT NUMBER AND NO DEVIATIONS WERE NOTED. THERE HAVE NOT BEEN ANY OTHER COMPLAINTS FOR THIS LOT NUMBER. DAMAGE APPEARS TO BE THE RESULT OF THE END USER, MAY HAVE OCCURED DURING UNPACKING OR DURING INSTALLATION TO THE APPLICATOR. NO CORRECTIVE/PREVENTIVE ACTION IS REQUIRED.

Description of Event or Problem

COUNTRY OF COMPLAINT: (B)(6). BROKEN CARTRIDGE; INCORRECT POSITIONING OF THE CLIP.