VERSACARE FRAME
Received Aug 20, 2015 · Event occurred Jul 28, 2015
Report 1824206-2015-00825 · MDR key 5019332
Device
Product problems
- Device Alarm System
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
THE HILL-ROM TECHNICIAN FOUND THE COMMUNICATION CABLE INOPERABLE. PER THE HILL-ROM USER MANUAL, WARNING: A COMMUNICATION CABLE MUST BE USED FOR BEDS THAT HAVE NURSE CALL. FAILURE TO DO SO COULD CAUSE PATIENT INJURY. FOR BEDS THAT HAVE NURSE CALL SYSTEMS, A COMMUNICATION CABLE MUST BE CONNECTED BETWEEN THE BED AND FACILITY COMMUNICATION SYSTEM. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE COMMUNICATION CABLE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem