inforMED
MalfunctionFRI

SCALE, STAND-ON, PATIENT

Received Aug 20, 2015 · Event occurred Feb 20, 2015

Report 5018057 · MDR key 5018057

Device

Generic name

Scale, Stand-on, Patient

Model number

H140-10-5

Product problems

  • Loss of Power
  • Smoking
  • Device Displays Incorrect Message
  • Sparking

Patient

44 YR

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

PATIENT WEIGHED BEFORE SURGERY. PATIENT'S WOOL SWEATER TOUCHED THE METAL HEIGHT BAR. THE RICE LAKE STANDING SCALE MADE A LOUD SPARK AND HAD SOME WHITE SMOKE. THEN THE RICE LAKE STANDING SCALE TURNED OFF. WHEN THE NURSE TURNED IT BACK ON IT HAD LINES "SLOP". THE PATIENT DID NOT SUSTAIN ANY INJURY.