VERSACARE FRAME

Received Aug 20, 2015 · Event occurred Jul 28, 2015

Report 1824206-2015-00828 · MDR key 5018046

Device

Generic name

A/c Powered Adjustable Hospital Bed

Model number

3200

Product problems

  • Device Slipped

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THE HILL-ROM TECHNICIAN FOUND THE BRAKES WERE NOT HOLDING DUE TO THE CASTERS CLICKING AND ROTATING WHILE THE BRAKES WERE NOT HOLDING IN BRAKE MODE. PER THE HILL-ROM SERVICE MANUAL, PERFORM ANNUAL PREVENTIVE MAINTENANCE PROCEDURES TO MAKE SURE ALL VERSACARE® BED COMPONENTS ARE FUNCTIONING AS ORIGINALLY DESIGNED. BRAKE CASTERS SHOULD BE CHECKED FOR CUTS, WEAR AND QUALITY OF TREAD, ETC. AND REPLACED WHEN NECESSARY. EXAMINE THE BRAKES TO SEE WHETHER THE BED MOVES WHEN THE BRAKE ARE SET. REPLACE AS NECESSARY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE CASTERS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).