inforMED
MalfunctionFNK

VERSACARE FRAME

Received Aug 20, 2015 · Event occurred Jul 28, 2015

Report 1824206-2015-00826 · MDR key 5018019

Device

Generic name

A/c Powered Adjustable Hospital Bed

Model number

3200

Product problems

  • Device Alarm System

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THE HILL-ROM TECHNICIAN FOUND THE COMMUNICATION CABLE INOPERABLE. PER THE HILL-ROM USER MANUAL, WARNING: A COMMUNICATION CABLE MUST BE USED FOR BEDS THAT HAVE NURSE CALL. FAILURE TO DO SO COULD CAUSE PATIENT INJURY. FOR BEDS THAT HAVE NURSE CALL SYSTEMS, A COMMUNICATION CABLE MUST BE CONNECTED BETWEEN THE BED AND FACILITY COMMUNICATION SYSTEM. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE COMMUNICATION CABLE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE NURSE CALL NOT WORKING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).