inforMED
MalfunctionIOY

SUPPORT ARM 177

Received Aug 13, 2015

Report 8010042-2015-00356 · MDR key 5015125

Add to watch list
Product code (IOY)
Manufacturer (Maquet Critical Care Ab)
Model (6481720)

Device

Generic name

Support, Arm

Catalog number

6481720

Product problems

  • Mechanical Problem

Patient

Not reported

  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

FURTHER INFO REGARDING THE EVENT HAS BEEN SOUGHT. A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED.

Description of Event or Problem

IT WAS REPORTED THAT THE SUPPORT ARM FOR THE VENTILATOR BROKE. THERE WAS NO PATIENT INVOLVEMENT. (B)(4).

Additional Manufacturer Narrative

(B)(4). ACCORDING TO INFORMATION RECEIVED, THE SUPPORT ARM BROKE AT THE JOINT ABOVE THE FIRST BEND. THE SUPPORT ARM IS BY DESIGN A CASTING PRODUCT. NO PARTS OR PICTURES WERE RECEIVED BACK FOR INVESTIGATION. NEITHER WAS INFORMATION OF HOW OR UNDER WHICH CIRCUMSTANCES THE SUPPORT ARM BROKE RECEIVED. PREVIOUS INVESTIGATIONS OF THE SUPPORT ARM LED TO A CHANGE OF THE MANUFACTURING PROCESS IN ORDER TO OBTAIN A HIGHER MECHANICAL STRENGTH OF THE SUPPORT ARM. THIS CHANGE WAS IMPLEMENTED IN PRODUCTION DURING SEPTEMBER 2009. IT IS UNKNOWN IF THE REPORTED SUPPORT ARM WAS MANUFACTURED BEFORE, OR AFTER, THE IMPLEMENTATION OF THIS CHANGE. THEREFORE, THE CAUSE FOR THIS EVENT CANNOT BE DETERMINED.

Description of Event or Problem

(B)(4). THIS FOLLOW-UP MEDWATCH WAS CREATED FROM A PREVIOUSLY RECEIVED INITIAL MEDWATCH WITH MANUFACTURER REPORT # : 8010042-2015-00356. THE ORIGINAL DATE OF THE INITIAL MEDWATCH REPORT WAS 07/24/2015. (B)(4).