inforMED
MalfunctionFNL

Centrella Bed

Received Jul 31, 2026 · Event occurred Jul 2, 2026

Report 1824206-2026-01752 · MDR key 50104834

Device

Generic name

Bed, Ac-powered Adjustable Hospital

Model number

P7900B

Catalog number

P7900B000011

Lot number

N/A

Product problems

  • No Audible Alarm

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

Baxter received a report stating that CENTRELLA MED-SURG bed exit was not alarming. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

Additional Manufacturer Narrative

The Baxter technician found the PCB needed to be calibrated. Per the Baxter Service Manual, it is necessary for the Centrella Bed to have an effective maintenance program. We recommend that you do annual preventative maintenance PM. Make sure the bed exit system operates correctly. Replace parts if necessary. A search of the Baxter maintenance records showed Baxter performed preventive maintenance on this bed (b)(6) 2025. It is unknown if the facility performed any other preventive maintenance on this bed. The technician calibrated PCB to resolve the reported event. Based on this information, no further action is required.