Centrella Bed
Received Jul 31, 2026 · Event occurred Jul 2, 2026
Report 1824206-2026-01752 · MDR key 50104834
Device
Generic name
Bed, Ac-powered Adjustable Hospital
Manufacturer
Baxter Healthcare CorporationModel number
P7900BCatalog number
P7900B000011
Lot number
N/A
Product problems
- No Audible Alarm
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
Baxter received a report stating that CENTRELLA MED-SURG bed exit was not alarming. There was no patient/user injury, medical intervention, symptom, or adverse event reported.
Additional Manufacturer Narrative
The Baxter technician found the PCB needed to be calibrated. Per the Baxter Service Manual, it is necessary for the Centrella Bed to have an effective maintenance program. We recommend that you do annual preventative maintenance PM. Make sure the bed exit system operates correctly. Replace parts if necessary. A search of the Baxter maintenance records showed Baxter performed preventive maintenance on this bed (b)(6) 2025. It is unknown if the facility performed any other preventive maintenance on this bed. The technician calibrated PCB to resolve the reported event. Based on this information, no further action is required.