Millennium M10 Concentrator

Received Jul 31, 2026 · Event occurred Jun 26, 2026

Report 2518422-2026-040891 · MDR key 50104314

Device

Generic name

Generator, Oxygen, Portable

Manufacturer

Respironics, Inc.

Model number

M10600

Catalog number

M10600

Product problems

  • Thermal Decomposition of Device

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

The manufacturer received a work order reporting that the Millennium M10 Concentrator device exhibited red light. During the evaluation of the device at the third-party service center, the device was visually inspected and thermal damage was detected.Additionally, the service technician also noted leakage and dirt.

Additional Manufacturer Narrative

The reported failure of thermal issue is accomadated in the product risk management file as being linked to Hazard with description Fire, flame, smoke. Complaints for this failure are reviewed via the Post Market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure. As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued. QA continues to monitor complaints for this issue per the cadence in the Complaint Trending procedures.