Endo Clip III
Received Jul 31, 2026 · Event occurred Jul 23, 2026
Report 9612501-2026-02093 · MDR key 50102299
Device
Generic name
Applier, Surgical, Clip
Manufacturer
Davis & Geck Caribe LtdModel number
176630BCatalog number
176630B
Lot number
J6E3729Y
Product problems
- Break
- Difficult to Remove
Patient
43 YR · Female
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
It was reported that during an emergency laparoscopic appendectomy, the clip applier passed smoothly through the port for use and clips were applied without any obvious issues. However, when the surgeon attempted to remove the clip applier back through the port, it became stuck. The surgeon used the laparoscope to examine the clip applier and observed a piece of metal sticking through the shaft of the applier. To remove the clip applier safely, the surgeon removed the port itself. There was no patient injury.
Additional Manufacturer Narrative
Medtronic submits this report to comply with FDA regulations 21 CFR Parts 4 and 803. Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report. This report does not constitute an admission or a conclusion by FDA, Medtronic, or its employees that the device, Medtronic, or its employee caused or contributed to the event described in the report. Any required fields that are unpopulated are blank because the information is currently unknown or unavailable. Medtronic will submit a supplemental report if additional relevant information becomes known.