inforMED
MalfunctionLWP

ACCOLADE MRI DR

Received Jul 31, 2026 · Event occurred Jul 15, 2026

Report 2124215-2026-41152 · MDR key 50094412

Device

Generic name

Pacemaker

Model number

L311

Catalog number

L311

Lot number

948974

Product problems

  • High impedance

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

It was reported that this pacemaker system triggered lead safety switch (LSS) due to a high out of range pacing impedance measurement of 2034 Ohms. This device remains in service. No adverse patient effects were reported.

Additional Manufacturer Narrative

This product investigation is still in progress. This report will be updated when product investigation is complete.