inforMED
MalfunctionLNS

STORZ MEDICAL

Received Nov 24, 2003 · Event occurred Nov 14, 2003

Report MW1030347 · MDR key 500906

Device

Generic name

Lithrotripsy Table

Manufacturer

Storz Medical

Catalog number

T-1-81052

Lot number

*

Product problems

  • Nonstandard Device
  • Defective Component
  • Electrical wires, defective

Patient

57 YR

  • Electric Shock
  • Surgical procedure aborted/stopped

Narrative

Description of Event or Problem

A LITHROTRIPSY WAS DISCOVERED TO HAVE A PROBLEM SO THE PROCEDURE WAS CANCELLED. A TECHNICIAN TOUCHED AN ANESTHESIA CART AND THE LITHROTRIPSY TABLE AT THE SAME TIME AND RECEIVED AN ELECTRICAL SHOCK. AN INVESTIGATION REVEALED THAT A VENDOR-MADE ADAPTOR CORD WAS IMPROPERLY WIRED (THE HOT (WHITE) WIRE WAS CONNECTED TO THE COMMON TERMINAL IN THE GROUND PLUG. THE GREEN/GROUND WIRE IN THE CORD WAS CONNECTED TO ONE OF THE 120V TERMINALS IN THE PLUG.