inforMED
MalfunctionHRX

DW ARTHROSCOPY FLUID MANAGEMENT DEV

Received Jul 30, 2026 · Event occurred Jul 13, 2026

Report 1220246-2026-04111 · MDR key 50083092

Device

Generic name

Arthroscope And Accessories

Manufacturer

Arthrex, Inc.

Catalog number

AR-6480

Lot number

32342712

Product problems

  • Increase in Pressure

Patient

Unknown

  • Joint Dislocation

Narrative

Description of Event or Problem

On 10-June-2026, it was reported by a Sales Representative via (b)(4) that an AR-6480 pump over pressured and expanded the shoulder joint. This occurred during use in a shoulder arthroscopy case with no patient effect.The pressure was set to 50. The pump shut off and an unknown alarm sounded. A different pump was used to complete the procedure. There was no revision to remove extra fluid and the patient was not kept at the facility for fluid removal.

Additional Manufacturer Narrative

Investigation is in process. A follow-up report will be provided upon availability of additional information.