inforMED
InjuryLIT

MUSTANG?

Received Jul 29, 2026 · Event occurred Jul 2, 2026

Report 2124215-2026-40573 · MDR key 50069737

Device

Generic name

Stents, Drains And Dilators For The Biliary Ducts

Model number

H74939171080870

Catalog number

H74939171080870

Lot number

0038732514

Product problems

  • Difficult to Remove
  • Material Rupture
  • Detachment of Device or Device Component

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

It was reported that balloon rupture and separation occurred, requiring additional intervention. The target lesion was located in an arteriovenous shunt. A 8.0 x 80, 75cm Mustang balloon catheter was advanced from the arterial side of the AV shunt through the loop to the venous side and across a previously implanted covered stent. During inflation at 18 atmospheres with a rated burst pressure of 20 atmospheres, the balloon ruptured and separated into two pieces. The balloon became entrapped by the covered stent and could not be withdrawn. Surgical intervention was performed to remove the retained balloon fragments. As a result of the event, the patients arteriovenous shunt could no longer be used for routine dialysis, and alternative dialysis access was required. No additional patient complications were reported as a result of this event.