inforMED
MalfunctionDSP

CARDIOSAVE HYBRID, 3.1 EDITION

Received Jul 29, 2026 · Event occurred Jul 8, 2026

Report 2249723-2026-0003642 · MDR key 50066508

Device

Generic name

System, Balloon, Intra-aortic And Control

Manufacturer

Datascope Corp.

Model number

0998-00-0800-65

Catalog number

0998-00-0800-65

Product problems

  • Mechanical Problem

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

It was reported that during clinical use, the CARDIOSAVE Intra-Aortic Balloon Pump (IABP) had a message indicating that maintenance needs to be considered has been generated. There was no patient harm or injury reported. A GETINGE Field Service Engineer (FSE) was dispatched to evaluate the IABP unit and was able to confirm from the Message log Fault # 78 (Exec UI Bridge Sender failed WDT), Fault # 79 (Video HeartBeat Lost) and #80 Video communication error. Running test OK (issue did not recur) and no abnormalities found in IABP. Work performed according to the service manual and Results were good.

Additional Manufacturer Narrative

E. Initial ReporterEvent Site Postal Code: (b)(6).