inforMED
InjuryFBO

Resectoscope Sheath, 24 Fr.

Received Jul 28, 2026 · Event occurred Jul 2, 2026

Report 9610617-2026-01748 · MDR key 50061012

Device

Generic name

Resectoscope Sheath, 24 Fr.

Model number

27242BZ

Catalog number

27242BZ

Lot number

TP01

Product problems

  • Detachment of Device or Device Component

Patient

Male

  • Insufficient Information
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

It was reported that during a TUR-B re-resection following the switch from the visualisation shaft to the resection loop, a metallic, cylindrical foreign body was detected in the bladder. It turned out to be the broken-off ceramic tip of the resectoscope shaft that had been used. The foreign body initially remained in the bladder and had to be removed after completion of the actual procedure using a Dormia basket. The retrieval was technically challenging but was carried out successfully and without causing trauma. According to current knowledge, no immediate harm to the patient occurred.

Additional Manufacturer Narrative

The device in question was returned to manufacturer as reflected in section D9. The investigation is not yet complete, investigation results are pending.The event is filed under internal KARL STORZ complaint ID: (b)(4).