inforMED
MalfunctionLIP

ARCHITECT Syphilis TP Reagent Kit

Received Jul 28, 2026 · Event occurred Jul 12, 2026

Report 3002809144-2026-00204 · MDR key 50057381

Device

Generic name

Enzyme Linked Immunoabsorption Assay, Treponema Pallidum

Manufacturer

Abbott Gmbh

Catalog number

08D06-87

Lot number

83318BE01

Product problems

  • False Negative Result

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

The customer observed false nonreactive ARCHITECT Syphilis TP results that do not match Antibio platform. The results provided were: Initial=0.55 S/CO (< 1.00 S/CO=nonreactive)/repeated=0.60 S/CO. Another hospital=positive.Autobio platform= positive. TPPA=positive; RPR=negative. There was no reported impact to patient management.

Additional Manufacturer Narrative

An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.Section A Patient Information: Refer to Section B5 for complete patient information. All available patient information was included. Additional patient details are not available.This report is being filed on an international product, list number 08D06-87, that has a similar product distributed in the US, list number 8D06-31 / 41, with 510K/PMA/BLA number K153730.