Received Jul 27, 2026 · Event occurred Jul 2, 2026

Report 2126677-2026-00006 · MDR key 50045316

Device

Generic name

System, X-ray, Stationary

Model number

XR656

Lot number

DX1SS1600127HL

Product problems

  • Device Difficult to Maintain

Patient

40 YR · Male

  • Laceration(s)

Narrative

Description of Event or Problem

On (b)(6) 2026, a GE HealthCare Field Engineer (FE) performing service activities on a Discovery XR656 fixed radiographic X-ray system located at (b)(6) Hospital - (b)(6) Saudi Arabia, reported an injury that occurred during the replacement of the overhead tube suspension (OTS) angulation brake. During the procedure, the X-ray tube assembly unexpectedly moved downward, entrapping the FE¿s hand and resulted in a 3 cm laceration. The injury required medical treatment, including the placement of four sutures.

Additional Manufacturer Narrative

GE HealthCare¿s investigation into the reported event has been initiated and is ongoing. A follow-up report will be submitted when the investigation has been completed.Legal Manufacturer: GE HANGWEI MEDICAL SYSTEMS CO., LTD. West Area of Building No.3, No.1 Yongchang North Road, Beijing Economic and Technological Development Area, Beijing, CHINA 100176.