inforMED
InjuryJCW

AMS 700 CX MS Pump

Received Jul 27, 2026 · Event occurred Jun 29, 2026

Report 2124215-2026-40043 · MDR key 50037256

Device

Generic name

Prosthesis, Penis, Inflatable

Model number

72404238

Catalog number

72404238

Lot number

1100875600

Product problems

  • Degraded
  • Material Puncture/Hole
  • Inadequacy of Device Shape and/or Size

Patient

Not reported

  • Urinary Retention

Narrative

Description of Event or Problem

It was reported that the patient with this Inflatable Penile Prosthesis (IPP) experienced urinary retention and required catheterization for several weeks. At the three week postoperative follow up visit, the physician noted buckling and breaks of the cylinders. Upon inspection, oversizing of the cylinders was suspected to have contributed to the buckling. As a result, the cylinders and pump were explanted and replaced. No further patient complications were reported.