inforMED
MalfunctionLRO

CENTURION Surgical Procedure Pak

Received Jul 27, 2026 · Event occurred Jul 2, 2026

Report 1644019-2026-04016 · MDR key 50033674

Device

Generic name

General Surgery Tray (kit)

Model number

NA

Catalog number

8065752200

Lot number

17UJ2P

Product problems

  • Incomplete or Inadequate Connection

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

A nurse reported that tip locker error was displayed during calibration for cataract surgery without intraocular lens implantation. The tip is tightened firmly several times, but the error message persisted. A new pk was opened and the error was disappeared. There was no patient contact and no patient impact.This report pertains to phacoemulsification tip involved in this reported event.

Additional Manufacturer Narrative

Investigation including root cause analysis is in progress. A supplemental MDR will be filed as necessary in accordance with 21 CFR 803.56 when additional reportable information becomes available. The manufacturer internal reference number is: (b)(4).