CENTURION Surgical Procedure Pak
Received Jul 27, 2026 · Event occurred Jul 2, 2026
Report 1644019-2026-04016 · MDR key 50033674
Device
Generic name
General Surgery Tray (kit)
Manufacturer
Alcon Research, Llc - HoustonModel number
NACatalog number
8065752200
Lot number
17UJ2P
Product problems
- Incomplete or Inadequate Connection
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
A nurse reported that tip locker error was displayed during calibration for cataract surgery without intraocular lens implantation. The tip is tightened firmly several times, but the error message persisted. A new pk was opened and the error was disappeared. There was no patient contact and no patient impact.This report pertains to phacoemulsification tip involved in this reported event.
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress. A supplemental MDR will be filed as necessary in accordance with 21 CFR 803.56 when additional reportable information becomes available. The manufacturer internal reference number is: (b)(4).