inforMED
MalfunctionMEB

INFUSOR

Received Jul 27, 2026 · Event occurred Jun 11, 2026

Report 1416980-2026-02441 · MDR key 50027608

Device

Generic name

Pump, Infusion, Elastomeric

Model number

J2C1711

Catalog number

J2C1711

Lot number

25M003

Product problems

  • No Flow

Patient

Male

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

It was reported that a Small Volume Infusor exhibited a slow flow rate, resulting in almost all or a large amount of the 110 mL solution remaining in the device during patient infusion. The device was filled with 46 mL of 5-Fluorouracil (5-FU) and 69 mL of saline having a total volume of 115 mL. The expected infusion duration was 46 hours. There was no report of patient injury or medical intervention associated with this event. No additional information was available.

Additional Manufacturer Narrative

Should additional relevant information become available, a supplemental report will be submitted.