inforMED
OtherHBF

CODMAN DISPOS PERFORATOR

Received Jul 27, 2026

Report 3014334038-2026-00059 · MDR key 50024259

Device

Generic name

Disposable Perforator

Catalog number

261221

Lot number

7550133

Product problems

  • Fail-Safe Did Not Operate

Patient

Not reported

  • Hemorrhage/Blood Loss/Bleeding

Narrative

Description of Event or Problem

A facility reported that when doing the second hole, the perforator (ID: 261221) did not disengage, causing a tear in the dura matter. The bleeding was controlled without difficulties, and the procedure was completed.- Is the motor electrically or pneumatically powered? "Electrically powered."- What is the brand of the motor used? "Medtronic."- Was the perforator properly assembled with the motor? "Yes."- Were the recommended tests performed between each hole? "Yes."- Was the approach angle perpendicular as specified in the Instructions for Use? "The approach angle was not strictly perpendicular". - Was constant downward pressure applied? Yes.

Additional Manufacturer Narrative

An investigation has been initiated based on the reported information. Upon completion of the investigation, a follow-up report will be submitted.