inforMED
MalfunctionHXX

Invictus Spinal Fixation System

Received Jul 27, 2026 · Event occurred Jun 30, 2026

Report 2027467-2026-00162 · MDR key 50008555

Device

Generic name

Screwdriver

Model number

17111

Catalog number

17111

Lot number

TI00031

Product problems

  • Break
  • Material Separation

Patient

Not reported

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

It was reported that the tip of the driver broke off during a case. The tip was retrieved.

Additional Manufacturer Narrative

Visual inspection confirms that the distal tip of driver has sheared off, along with a portion of the distal shaft. The failure is likely due to input torque exceeding the strength of the tip. Review of device history records showed no manufacturing or processing related irregularities. Labeling review:WARNINGS/CAUTIONS/PRECAUTIONS: Risks identified with the use of these devices, which may require additional surgery, include device component failure, loss of fixation/stabilization, non-union, vertebral fracture, neurological injury, vascular or visceral injury. POSSIBLE ADVERSE EFFECTS: Initial or delayed loosening, disassembly, bending, dislocation, and/or breakage of device components.