inforMED
InjuryJWH

GMK Sphere Total Knee System

Received Jul 27, 2026 · Event occurred Jul 13, 2026

Report 3005180920-2026-00778 · MDR key 50005868

Device

Generic name

Gmk-sphere Tibial Insert E-cross - Flex 3l - 11mm

Model number

02.12.E0311FL

Lot number

2002475

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

74 YR · Female

  • Joint Laxity

Narrative

Description of Event or Problem

At about 1 year 11 months after the primary, the patient came in presenting knee instability. The surgeon revised the insert 11mm to 14mm insert for stability. The surgery was completed successfully. Several months prior to this revision, the patient fractured her ankle and dislocated her patella.

Additional Manufacturer Narrative

Batch review performed on 20 July 2026 Lot 2002475: (b)(4) items manufactured and released on 14-July-2020. Expiration date: 2025-07-02. No anomalies found related to the problem. To date, all items of the same lot have been sold with no similar reported events during the period of review. Root cause: The surgeon revised liner height, which is a common practice to restore the joint stability. There is no indication that any potential issue with the device may have caused or contributed to the event.