inforMED
InjuryLZO

SMS Stem

Received Jul 27, 2026 · Event occurred Jul 17, 2026

Report 3005180920-2026-00791 · MDR key 50005866

Device

Generic name

Sms Collared Solid Stem Std Size9

Model number

01.36.149

Lot number

2209145

Product problems

  • Osseointegration Problem

Patient

Unknown

  • Inadequate Osseointegration
  • Implant Pain
  • Aseptic loosening

Narrative

Description of Event or Problem

Revision surgery due to stem loosening after about 3 years after primary. The surgeon revised the Medacta stem with a competitor stem and revised Medacta head and liner with a new Medacta head and liner.

Additional Manufacturer Narrative

Batch review performed on 23 July 2026: Lot 2209145: (b)(4) items manufactured and released on 28-Sep-2022. Expiration date: 19-Sep-2027. No anomalies found related to the problem. To date, (b)(4) the items of the same lot have been sold with no similar reported event during the period of review. Root cause: