inforMED
InjuryKRO

GMK Hinge Total Knee System

Received Jul 27, 2026 · Event occurred Jul 16, 2026

Report 3005180920-2026-00780 · MDR key 50005842

Device

Generic name

Gmk-hinge Tibial Insert - Size1 - 17 Mm

Model number

02.09.0117H

Lot number

2114858

Product problems

  • Device Appears to Trigger Rejection

Patient

Not reported

  • Medical device site infection

Narrative

Description of Event or Problem

Revision surgery after 1 year and 11 months from the Medacta primary due to infection. The surgeon revised the insert to the same size insert.

Additional Manufacturer Narrative

Batch review performed on 21 July 2026 02.09.0117H GMK-HINGE Tibial Insert - Size1 - 17 mm (K130299) Lot. 2114858: (b)(4) items manufactured and released on 01 February 2022. Expiration date: 18 January 2027. No anomalies found related to the problem. To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review. Root cause: Infection is a known possible complication of any surgery. Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.