inforMED
InjuryMSS

EnVista SimplifEYE Injector

Received Jul 27, 2026 · Event occurred Jun 25, 2026

Report 0001313525-2026-00106 · MDR key 50005681

Device

Generic name

Manual Intraocular Lens Injector, Single-use

Manufacturer

Bausch + Lomb Inc.

Model number

21987

Lot number

2304H10

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

It was reported that during insertion of an intraocular lens (IOL) the lens became stuck and would not implant fully. Unable to remove the lens from the incision, it was enlarged approximately 1.2mm with a keratome and a back up lens was attempted to be implanted. The back up lens became stuck and a Drysdale instrument was used to remove the lens from the injector. However, zonnuals were destroyed. It has was placed in the center of the capsule bag and it remains implanted. Surgery was delayed 10 minutes. This report is for the second injector, 4 of 4. Related report number: 0001313525-2026-00104, 0001313525-2026-00107 and 0001313525-2026-00105.

Additional Manufacturer Narrative

The device is not available for evaluation. The investigation of this event is in progress. A follow-up report will be submitted upon completion of investigation.