inforMED
InjuryKNT

CARDINAL HEALTH

Received Jul 27, 2026 · Event occurred May 13, 2026

Report 1423537-2026-00241 · MDR key 50000546

Device

Generic name

Tubes, Gastrointestinal (and Accessories)

Model number

8884711006E

Catalog number

8884711006E

Product problems

  • Improper or Incorrect Procedure or Method

Patient

Not reported

  • Dehydration
  • Hyponatremia

Narrative

Description of Event or Problem

The customer reported the patient was found to have a Dobhoff feeding tube with the stylet still in place several days after placement which led to dehydration and hyponatremia. Per the customer, while the metal stylet is essential for placement of the tube, removal of the stylet is a challenge. The stylet plastic tip is the very same color as the ports on the feeding tube and can be difficult to differentiate easily for users. There are two ports, so inexperienced users may think the stylet is supposed to be there and use the smaller port, restricting the flow rate of feeds and fluids. Per additional information provided on July 21, 2026, the treatment for dehydration was IV fluids x 24hours. There was no treatment provided for the hyponatremia which resolved prior to discharge, monitoring only. The tube was in place for 5 days before the stylet was found to still be in the tube. The stylet was removed and the tube remained in place with placement confirmed by x-ray.