inforMED
InjuryMBP

XEMFLIFI DBM PLUS, 3CC

Received Aug 7, 2015 · Event occurred Mar 13, 2015

Report 3004142400-2015-00032 · MDR key 4990515

Device

Generic name

Xemplifi Dbm

Model number

8103.0203S

Lot number

1362376553

Product problems

  • Insufficient Device Problem Information

Patient

50 YR

  • Abscess
  • Unspecified Infection

Narrative

Additional Manufacturer Narrative

A COMPREHENSIVE INVESTIGATION WAS IMMEDIATELY INITIATED ON RECEIPT OF THE COMPLAINT. THE PRODUCT WAS NOT RETURNED FOR EVALUATION. A THOROUGH INVESTIGATION WAS COMPLETED, AND NO DEVIATIONS TO PROCESSING WERE IDENTIFIED. THERE HAVE BEEN NO PREVIOUS REPORTED ISSUES OF THIS TYPE. XEMPLIFI PLUS IS A STERILE PACKAGED PRODUCT. STERILITY CAN BE ASSURED AS LONG AS THE PRODUCT PACKAGE IS NOT COMPROMISED. IN THIS CASE, THERE WAS NO REPORTED ISSUES WITH THE STERILITY OF THE PRODUCT. POSTOPERATIVE INFECTION CAN BE CAUSED FROM A WIDE VARIETY OF SOURCES. IN THIS EVENT, IT IS NOT POSSIBLE TO DETERMINE THE EXACT CAUSE OF THE INFECTION. THIS REPORT IS ASSOCIATED WITH MDR (B)(4).

Description of Event or Problem

IT WAS REPORTED TO GLOBUS THAT A PATIENT HAD A FUSION SURGERY C5-6 ON (B)(6) 2016 AT (B)(6) HOSPITAL (B)(6). THE PATIENT LATER DEVELOPED AN ABSCESS/INFECTION, AND HAD TO HAVE THE SPINAL HARDWARE, COALITION SPACER, AND XEMPLIFI PUTTY REMOVED. THE REMOVAL SURGERY TOOK PLACE (B)(6) 2015. THE REMOVAL SURGERY WAS PERFORMED AT (B)(6) HOSPITAL (B)(6). THE REMOVAL SURGERY WAS PERFORMED BY DIFFERENT SURGEON THAN THE INITIAL SURGERY.