HEART-LUNG MACHINE HL20

Received Aug 3, 2015 · Event occurred Jul 24, 2015

Report 8010762-2015-00866 · MDR key 4972743

Device

Model number

70102.8580

Catalog number

70102.8580

Product problems

  • Device Displays Incorrect Message

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

A FIELD SAFETY TECHNICIAN OPENED THE RPM AND ONE BELT IS BROKEN (THE SECOND ONE SEEMS TO BE OK). HE MEASURED THE BELT TENSION 445 HZ. POSSIBLE ROOT CAUSE: DEFECTIVE BELT FROM A FACTORY, TENSION TOO HIGH, MAYBE COMBINATION OF BOTH. HE REPAIRED (BELT REPLACEMENT) / ORDERING NEW MOTOR ISOLATION KIT FOR BETTER ADJUSTMENT OF TENSION. THE DATA IS ALSO BEING HANDLED THROUGH A DESIGNATED MAQUET CARDIOPULMONARY TRENDING AND APPLICABLE INVESTIGATION PROCESS. DUE TO THIS NO FURTHER INVESTIGATION INITIATIONS WILL BE COMPLETED AT THIS TIME.

Description of Event or Problem

IT WAS REPORTED THAT DURING THE SURGERY, THE RPM MODULE REPORTED AN ERROR "BELT SLIP". THE PERFUSIONIST STOPPED THE PUMP BUT THE ERROR WAS STILL THERE. SO THEY STOPPED THE "SUCTION" FUNCTION. (B)(4).