TRIAGE PROFILER SOB PANEL

Received Jul 30, 2015 · Event occurred Jul 1, 2015

Report 2027969-2015-00552 · MDR key 4955101

Device

Generic name

Cardiac Marker Test

Model number

97300EU

Lot number

W59664

Product problems

  • Incorrect Or Inadequate Test Results

Patient

Not reported

  • Thrombosis

Narrative

Additional Manufacturer Narrative

INVESTIGATION CONCLUSION: NO DISCREPANT LOW RESULTS WERE OBSERVED FROM ANY DEVICES TESTED WITH CAL H. DATA RESULTS FROM ALL DEVICES TESTED WITH CAL H WERE WITHIN 2SD OF THE EXPECTED VALUE AND A CV OF 5.58% IS WITHIN MANUFACTURING FINAL RELEASE SPECIFICATIONS. REVIEWED THE MANUFACTURING BATCH RECORDS FOR SOB LOT W59664. NO ISSUES WITH D-DIMER RECOVERY OBSERVED. NO NCS OR TIFS WERE GENERATED ON THE LOT DURING PRODUCTION. LOT PASSED ALL FINAL RELEASE SPECIFICATIONS. AS OF (B)(6) 2015, THIS IS THE ONLY DISCREPANT LOW COMPLAINT AGAINST SOB LOT W59664. NO SAMPLE WAS RETURNED. UNABLE TO RULE OUT SAMPLE SPECIFIC INTERFERENCE AS A POTENTIAL CAUSE FOR DISCREPANT RESULTS. NO PRODUCT DEFICIENCY WAS ESTABLISHED WITH THE TRIAGE SOB DEVICE LOT. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.

Description of Event or Problem

DISCREPANT SOB PROFILER RESULTS. EVENT OCCURRED IN (B)(6). PATIENT CAME TO PHYSICIAN'S OFFICE WITH ACUTE DEEP VEIN THROMBOSIS. DOCTOR GAVE HEPARIN AND SENT PATIENT TO CARDIOLOGIST. CARDIOLOGIST MEASURED A DDIMER OF 250 WITH SOB PROFILER. SECOND MEASUREMENT 260 NG/ML. TIME BETWEEN MEASUREMENTS NOT REPORTED. THROMBOSIS WAS VERIFIED BY ULTRASOUND. NO MEDICATION OF THE PATIENT. PATIENT WAS ASKED TO RETURN TO PHYSICIAN'S OFFICE IN 3 MONTH. ALTHOUGH REQUESTED, NO ADDITIONAL INFORMATION PROVIDED. NO REPORTED ADVERSE EVENTS.