inforMED
MalfunctionDPW

HL-20 TWIN PUMP

Received Jul 28, 2015 · Event occurred Oct 11, 2013

Report 8010762-2014-00921 · MDR key 4944785

Device

Generic name

Flowmeter, Blood, Cardiovascular

Model number

70102.8692

Catalog number

MCP00703277

Product problems

  • Electrical /Electronic Property Problem
  • Device Displays Incorrect Message

Patient

Not reported

  • No Patient Involvement

Narrative

Description of Event or Problem

ON (B)(6) 2013 AT THE (B)(6), IT WAS REPORTED THROUGH THE MSUPPORT SYSTEM COMPLAINT NUMBER (B)(4) THAT THE MOTOR CONTROL BOARD AND CONTROL BOARD ON THE TPM 20-330 TWIN ROLLER PUMP MODULE SERIAL NUMBER (B)(4) WERE DEFECTIVE AFTER THE UNIT WAS UPGRADED TO SOFTWARE VERSION 2.5. THE SSU REPORTED THAT THE UNIT DISPLAYED AN A/D CONVERTER (+5V ERROR). (B)(4).

Additional Manufacturer Narrative

MAQUET MEDICAL SYSTEM, USA SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET CARDIOPULMONARY (B)(4). THE DEVICE WAS EVALUATED BY A SSU TECHNICIAN AND THE MOTOR CONTROL BOARD AND CONTROL BOARD WERE FOUND TO BE DEFECTIVE AND WERE REPLACED (B)(4).