HL-20 TWIN PUMP
Received Jul 28, 2015 · Event occurred Sep 5, 2013
Report 8010762-2014-00914 · MDR key 4944784
Device
Generic name
Flowmeter, Blood, Cardiovascular
Manufacturer
Maquet Cardiopulmonary AgProduct problems
- Device Displays Incorrect Message
Patient
Not reported
- No Information
Narrative
Description of Event or Problem
ON (B)(6) 2013 IT WAS REPORTED BY THE (B)(4) AT MAQUET MEDICAL (B)(4) THROUGH THE MSUPPORT SYSTEM COMPLAINT NUMBER (B)(4) THAT THE TPM 20-330 TWIN ROLLER PUMP MODULE SERIAL NUMBER (B)(4) DISPLAYED A BELTSLIP ERROR WHEN RUNNING AT 250 RPM BUT THE DISPLAY INDICATED 119 RPM. THE RPM WAS MEASURED WITH AN EXTERNAL TACHOMETER AND IT INDICATED 119 RPM.
Additional Manufacturer Narrative
MAQUET MEDICAL SYSTEM, USA SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET CARDIOPULMONARY (B)(4). THE DEVICE WAS EVALUATED BY A (B)(4) TECHNICIAN AND THE OPTICAL TACHO BOARD, MOTOR BELT AND BELT PULLEY WERE REPLACED AND THE ISSUE WAS RESOLVED. (B)(4).