MalfunctionDPW

HL-20 TWIN PUMP

Received Jul 28, 2015 · Event occurred Dec 19, 2013

Report 8010762-2014-00975 · MDR key 4944755

Device

Generic name

Flowmeter, Blood, Cardiovascular

Model number

70103,5075

Catalog number

MCP00703323

Product problems

  • Device Displays Incorrect Message

Patient

Not reported

  • No Patient Involvement

Narrative

Description of Event or Problem

ON (B)(6) 2013, AT THE (B)(6) HOSPITAL, (B)(6), IT WAS REPORTED THAT THE TPM 20-330 TWIN ROLLER PUMP MODULE SERIAL NUMBER (B)(4) DISPLAYED A HEAD ERROR ON THE RIGHT SIDE PUMP WHEN POWERED ON. THE DEVICE WAS REMOVED FROM SERVICE AND THE CUSTOMER WAS GIVEN A LOANER UNIT. THE NON-CONFORMITY REPORT INDICATES THAT THE FAILURE OCCURRED PRIOR TO USE ON A PATIENT. (B)(4).

Additional Manufacturer Narrative

MAQUET MEDICAL SYSTEM, USA SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET CARDIOPULMONARY (B)(4). THE UNIT WAS INVESTIGATED BY A SSU SERVICE TECHNICIAN AND THE OPTICAL TACHOMETER PCB WAS FOUND TO BE DEFECTIVE AND WAS REPLACED. (B)(4).