inforMED
MalfunctionLBI

CENTURION EXCEL

Received Jul 22, 2015 · Event occurred Jun 12, 2012

Report 2028295-2015-00001 · MDR key 4943577

Device

Generic name

Minor Surgical Light

Model number

061514

Product problems

  • Break
  • Off-Label Use

Patient

Not reported

  • Injury

Narrative

Additional Manufacturer Narrative

THE LIGHT AND ARM WERE RETURNED TO THE FACTORY FOR ANALYSIS. UPON ARRIVAL, THE ARM SUPPORT FITTING WAS INSPECTED FOR POROSITY AND/OR VOIDS IN THE MATERIAL. NO ANOMALIES IN THE MATERIAL WERE FOUND. THE ARM SUPPORT MATERIAL WAS ALSO IDENTIFIED THAT IT WAS THE SPECIFIED MATERIAL. ADDITIONAL TESTING OF NEW PRODUCT WAS PERFORMED TO VERIFY PRODUCTION PARTS PERFORMED AS DESIGNED. UNABLE TO DETERMINE EXACT CAUSE OF THIS FAILURE. WILL MONITOR FOR ANY FUTURE OCCURRENCES. THE ARM SUPPORT COMPONENT WAS PUT UNDER TEST WITH A 6X LOAD TO CONFIRM THE INITIAL LOAD HANDLING CAPACITY. THIS TEST WAS CONFIRMED AS POSITIVE AND THE CASE WAS CLOSE WITH THE REPLACEMENT OF THE LIGHT ON (B)(4) 2012.

Description of Event or Problem

ON (B)(6) 2012 AN INCIDENT WAS REPORTED TO MEDICAL ILLUMINATION REGARDING AN EXCEL LIGHT ARM SUPPORT FITTING FRACTURED CAUSING THE LIGHT TO FALL HITTING THE DOCTOR ON HIS BACK. THE LIGHT FELL WHEN BEING REMOVED.