inforMED
MalfunctionJIO

COMBUR 10 TEST

Received Jul 27, 2015 · Event occurred Jul 2, 2015

Report 1823260-2015-03866 · MDR key 4943173

Device

Generic name

Blood, Occult, Colorimetric, In Urine

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

04510062171

Lot number

22933001

Product problems

  • False Negative Result

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

ALL TEST STRIP SPECIFICATIONS HAVE BEEN MET FOR THE RETENTION SAMPLES TESTED. THE CUSTOMER HAS NOT MADE ANY FURTHER COMPLAINTS ABOUT THE STRIP LOT USED.

Description of Event or Problem

THE CUSTOMER COMPLAINED OF FALSE NEGATIVE ERYTHROCYTES (ERY) RESULTS FOR AN UNKNOWN NUMBER OF PATIENTS WHEN TESTED WITH "STUDY" COMBUR-10 URINE TEST STRIPS. THE CUSTOMER IS COMPARING THE COMBUR-10 URINE STRIPS TO SIEMENS MULTISTIX WHICH ARE TESTED ON THE SIEMENS CLINITEK 50 INSTRUMENT. THE ACTUAL RESULTS HAVE NOT BEEN PROVIDED. THIS INFORMATION HAS BEEN REQUESTED. THE DATE OF EVENT IS NOT KNOWN. IT IS NOT KNOWN IF THE RESULTS ARE BEING USED FOR DIAGNOSTIC PURPOSES. THIS INFORMATION HAS BEEN REQUESTED. IT IS NOT KNOWN IF ERRONEOUS RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. IT IS NOT KNOWN IF AN ADVERSE EVENT OCCURRED. NO ADVERSE EVENT WAS REPORTED. WHILE THE COMBUR-10 URINE STRIPS ARE NOT SOLD IN THE US, THE ROCHE URINE CHEMSTRIP IS SIMILAR. RETENTION SAMPLES OF THE SAME LOT THAT THE CUSTOMER USED WERE TESTED. THE REACTION COLORS WERE ACCEPTABLE AND DID NOT SHOW ANY ABNORMALITIES. NO FALSE NEGATIVE RESULTS WERE OBSERVED.

Additional Manufacturer Narrative

THIS EVENT OCCURRED IN (B)(6).