inforMED
MalfunctionMOU

BETA - CATH UNIT

Received Oct 17, 2003 · Event occurred Sep 18, 2003

Report MW1029890 · MDR key 492833

Device

Generic name

Brachytherapy Catheter

Manufacturer

*

Model number

*

Catalog number

*

Lot number

*

Product problems

  • Difficult To Position

Patient

82 YR

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

MD FELT CATHETER NEEDED TO BE REPOSITIONED SLIGHTLY & SOURCES RETURNED TO THE BETA-CATH APPLICATOR. SOURCES WERE NOT RETURNED WITHIN 5 SECONDS. MD HAD REPOSITIONED CATHETER & THE SOURCES WERE RETURNED TO THE STENT & IRRADIATION OF THE VESSELS WAS COMPLETED W/DIFFICULTY. FULL ASSESSMENT OF CATHETER MADE BY APPROPRIATE PERSONNEL TO DETERMINE CAUSE OF DYSFUNCTION. PROCEDURE COMPLETED W/O DIFFICULTY & NO EXCESSIVE EXPOSURE TO PT OR STAFF.