inforMED
OtherLJC

CHLAMYDIAZYME EIA

Received Oct 9, 2003 · Event occurred Sep 10, 2003

Report 1415939-2003-00012 · MDR key 489406

Device

Generic name

Eia For The Detection Of Chlamydia Antigen

Model number

NA

Catalog number

1803-30

Lot number

04572M100

Product problems

  • False Negative Result

Patient

UNKNOWN

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

THE ACCOUNT GENERATED NEGATIVE CHLAMYDIAZYME EIA RESULTS, REAGENT LOT 04572M100, ON 18 PTS THAT TESTED BSA SLIDE TECHNIQUE POSITIVE. THIS REPORT IS FOR PT 4 OF THE 18 TOTAL PTS. ALTHOUGH THE CONTROL VALUES WERE ACCEPTABLE, THE ACCOUNT NOTICED A BIG DROP IN THE POSITIVE RESULT VALUES AS COMPARED TO THE USUAL OBSERVED POSITIVE RESULT VALUES. THE ACCOUNT RETESTED THE PTS USING CHLAMYDIAZYME EIA REAGENT LOT 03228M100 WITH THE EXPECTED POSITIVE RESULTS. AMENDED REPORTS WERE SUBMITTED TO THE PHYSICIANS. IT IS UNKNOWN IF THERE WAS IMPACT TO PT MANAGEMENT. NO PT INFO IS AVAILABLE.