inforMED
InjuryCDX

BECKMAN COULTER ACCESS TOTAL T3

Received Sep 18, 2003 · Event occurred Aug 1, 2003

Report MW1029611 · MDR key 488463

Device

Generic name

Reagent

Manufacturer

Beckman Coulter

Model number

*

Catalog number

*

Lot number

215114

Product problems

  • High Test Results

Patient

*

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

CHEMISTRY INSTRUMENT FOR TOTAL T3 TESTING CHANGED IN JULY 2003 AFTER EXTENSIVE VALIDATION TESTS. IN AUGUST REVIEW OF PT TESTING SHOWED HIGHER THAN NORMAL NUMBER OF ELEVATED TOTAL T3 LEVELS. RETESTING AND INVESTIGATION REVEALED THAT THE CO CHANGED ITS REAGENT WITH NO LEAD TIME NOTICE PROVIDED TO REPORTER AT THE TIME. SAMPLES FROM ORIGINAL VALIDATION WERE RETESTED BY ANOTHER HOSP WHICH USES SAME REAGENT AND MANUFACTURER AND SAME HIGHER RESULTS WERE DEMONSTRATED. REPORTER ALSO SENT TO INDEPENDENT LAB AT THEIR EXPENSE; TESTS SHOWED APPROPRIATE TOTAL T3 LEVELS. REPORTER IS NOW SENDING TOTAL T3 TO ANOTHER LAB UNTIL THE PROBLEM CAN BE CORRECTED. CO CONTACTED- THEY RETESTED VALIDATION SAMPLES AND FOUND DISCREPANCIES AS WELL.