SINGLE USE 3-LUMEN SPHINCTEROTOME

Received Jun 11, 2015 · Event occurred May 7, 2015

Report 8010047-2015-00430 · MDR key 4845349

Device

Generic name

Sphincterotome

Model number

KD-V411M-0720

Lot number

K5220

Product problems

  • Break

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

DURING THE INVESTIGATION, OMSC FOUND THE PLASTIC COATING ON THE CUTTING WIRE WAS PARTIALLY MISSING. THERE WAS NO PT INJURY REGARDING THIS EVENT. THE INVESTIGATION CONFIRMED THAT THE CUTTING WIRE WAS BROKEN AT THE COATED PORTION AND THE BROKEN SECTION WAS MELTED AND BURNED. APPROXIMATELY COATING WAS MISSING FOR 5MM FROM THE BROKEN POINT. THERE WERE NOT OTHER ABNORMALITIES RELATED TO THE BREAKAGE IN THE SUBJECT DEVICE. ALSO AS THE CHECKING OF THE MANUFACTURING RECORD OF THE SAME LOT, NOTHING ABNORMAL WAS DETECTED. AS THE RESULTS OF THE INVESTIGATION, OMSC ASSUMES THAT THE DAMAGE OF THE COATING OCCURRED DUE TO CONTACTING WITH THE METAL PART OF THE FORCEPS ELEVATOR OF THE ENDOSCOPE. THE EXPOSED CUTTING WIRE FROM THE DAMAGED COATING CONTACTED OR CAME CLOSE TO THE METAL PART OF THE FORCEPS ELEVATOR WHILE ACTIVATING THE OUTPUT, WHICH CAUSED SPARK AND A PART OF THE CUTTING WIRE BECAME EXTREMELY HOT, RESULTING IN BREAKAGE. THE COATING BROKE OFF AND CAME OFF FROM THE CUTTING WIRE BECAUSE OF THE USER HANDLING AFTER THE CUTTING WIRE WAS BROKEN. THE DEVICE MANUAL HAS WARNED USERS THAT "WHEN INSERTING THE INSTRUMENT INTO THE ENDOSCOPE, BE SURE THAT THE CUTTING WIRE IS PARALLEL TO THE TUBE. OTHERWISE, THE METAL PART OF THE FORCEPS ELEVATOR MAY CONTACT THE CUTTING WIRE AND PEEL OFF THE COATING MATERIAL." THIS REPORT IS BEING SUBMITTED AS MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.

Description of Event or Problem

OLYMPUS MEDICAL SYSTEMS CORP. (OSMC) WAS INFORMED THAT DURING ENDOSCOPIC SPHINCTEROTOMY (EST), THE CUTTING KNIFE BROKE. THE DOCTOR COMPLETED THE PROCEDURE BY USING ANOTHER DEVICE. .