SONICATOR 740X
Received Jun 5, 2015 · Event occurred Apr 7, 2015
Report 4837611 · MDR key 4837611
Device
Generic name
Stimulator, Ultrasound And Muscle, For Use In Applying Thera
Manufacturer
Mettler Electronics Corp.Model number
ME740Catalog number
*
Lot number
*
Product problems
- Output above Specifications
Patient
*
- Shock
Narrative
Description of Event or Problem
WHILE A STUDENT PHYSICAL THERAPIST WAS ADMINISTERING AN ULTRASOUND TREATMENT, THE PATIENT STATED THEY FELT AN ELECTRICAL SHOCK. PATIENT WAS NOT HARMED ONLY STARTLED. THE UNIT WAS REMOVED FROM SERVICE AND SENT TO THE CLINICAL ENGINEERING DEPT.======================MANUFACTURER RESPONSE FOR ULTRASOUND THERAPY UNIT, METTLER (PER SITE REPORTER).======================TECH SUPPORT STATED THAT MOST LIKELY THERE WAS A SHORT IN THE CABLE THAT CONNECTS TO THE PROBE CAUSING AN INCREASE IN THE OUTPUT...NOT AN ELECTRICAL SHOCK.