inforMED
OtherMOU

GALILEO INTRAVASCULAR RADIOTHERAPY CARTRIDGE, SOURCE WIRE

Received Aug 22, 2003 · Event occurred Mar 25, 2003

Report 1651662-2003-00013 · MDR key 483419

Device

Generic name

Radiation System Cartridge

Model number

1008112-0C

Catalog number

NA

Lot number

NA

Product problems

  • Method, improper/incorrect
  • Difficult To Position

Patient

UNKNOWN

  • Radiation Overdose

Narrative

Description of Event or Problem

DURING A PATIENT TREATMENT USING THE GALILEO SYSTEM, IT WAS REPORTED THAT THE ACTIVE WIRE WAS PLACED 6MM PROXIMAL TO THE DISTAL CATHETER MARKER AND DWELLED FOR 17 SECONDS. THE PHYSICIAN STOPPED THE PROCEDURE DUE TO THE IMPROPER PLACEMENT OF THE WIRE. THE WIRE WAS THEN REPOSITIONED AND DWELLED FOR 13 SECONDS BEFORE BEING STOPPED AGAIN FOR IMPROPER ACTIVE WIRE PLACEMENT. THE WIRE WAS POSITIONED A THIRD TIME AND A FULL TREATMENT WAS DELIVERED WITH A DWELL OF 144 SECONDS. THE DOSE DELIVERED TO THE PORTION INCLUDED WITH THE FIRST TWO DWELLS WAS REPORTED TO BE 2833CGY.

Remedial action

  • Other
  • Relabeling