GALILEO INTRAVASCULAR RADIOTHERAPY CARTRIDGE, SOURCE WIRE

Received Aug 22, 2003 · Event occurred Jun 20, 2002

Report 1651662-2003-00010 · MDR key 483415

Device

Generic name

Radiation System Cartridge

Model number

1008114-0C

Catalog number

NA

Lot number

NA

Product problems

  • Displacement
  • Method, improper/incorrect

Patient

UNKNOWN

  • Radiation Overdose

Narrative

Description of Event or Problem

THE SITE WAS PERFORMING A TANDEM TREATMENT. THE PROXIMAL REFERENCE LUMEN DIAMETER (RLD) WAS DETERMINED TO BE 3.3 MM AND THE DISTAL RLD WAS DETERMINED TO BE 2.8 MM. THE PROXIMAL SEGMENT WAS TREATED SUCCESSFULLY. THE RLD VALUE FOR THE DISTAL SEGMENT WAS MISTAKENLY ENTERED AS 3.3 MM RATHER THAN 2.8 MM AS WAS MEASURED. THE SITE INVESTIGATED THE POSSIBLE MISADMINISTRATION WITH THE DEPTH DOSE TABLES PROVIDED IN THE GALILEO INSTRUCTIONS FOR USE. THE SITE CALCULATED THE DOSE DELIVERED TO BE 26.4GY.

Remedial action

  • Other