VIDAS LYME IGM
Received Jun 5, 2015 · Event occurred Nov 1, 2014
Report 3002769706-2015-00038 · MDR key 4820698
Device
Generic name
Reagent, Borrelia Serological Reagent
Manufacturer
Biomerieux, S.a.Catalog number
30319
Lot number
1003354530
Product problems
- False Positive Result
Patient
Not reported
- No Consequences Or Impact To Patient
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
A CUSTOMER CONTACTED BIOMERIEUX REGARDING A FALSE POSITIVE RESULT ASSOCIATED WITH A NATIONAL QUALITY CONTROL (CNQ) EXERCISE PERFORMED IN (B)(6) 2014. THE IGM RESULT FOR THIS SAMPLE WAS EXPECTED TO BE NEGATIVE. THE VIDAS LYME IGM TEST, REF. (B)(4) GAVE A POSITIVE RESULT WHILE THE REAGENTS FOR OTHER COMPANIES GAVE THE EXPECTED NEGATIVE RESULT. CNQ IS AN EXTERNAL QUALITY CONTROL SCHEME ORGANIZED BY (B)(6) ON AN ANNUAL BASIS. IT CONSISTS OF SEVERAL BLIND SAMPLES WHICH ARE TESTED BY DIFFERENT CUSTOMERS ON THE SAME TECHNIQUE WITH THE AIM OF COMPARING THE RESULTS AND VALIDATE THE TECHNIQUE. THE L3 SAMPLE OF THE NATIONAL QUALITY CONTROL (CNQ) EXERCISE IS COMPOSED OF LYOPHILIZED PLASMA OF ONE PATIENT. THE VIDAS LYME IGM (LYM) ASSAY IS AN AUTOMATED QUALITATIVE TEST INTENDED FOR USE ON THE VIDAS FAMILY OF INSTRUMENTS, FOR THE DETECTION OF IGM ANTIBODIES TO BORRELIA BURGDORFERI SENSU LATO (SL) IN HUMAN SERUM OR PLASMA, TO AID IN THE DIAGNOSIS OF LYME DISEASE. THE REPORTED VIDAS LYME IGM, REF. (B)(4), IS NOT SOLD OR DISTRIBUTED WITHIN THE UNITED STATES; HOWEVER, A SIMILAR VIDAS LYME IGM ASSAY, REF (B)(4), IS SOLD AND DISTRIBUTED WITHIN THE UNITED STATES. AN INVESTIGATION WAS INITIATED AND THE FALSE POSITIVE RESULT WAS CONFIRMED WHEN TESTED WITH 2 DIFFERENT LOT NUMBERS OF THE REAGENT DAS LYME IGM. AN INTERFERENCE BETWEEN THE REAGENT AND THE SAMPLE L3 IS SUSPECTED. A SUBSEQUENT CAPA HAS BEEN INITIATED TO DETERMINE THE EXACT ROOT CAUSE AND PROPOSE CORRECTIVE / PREVENTIVE ACTIONS. THE CAPA IS ONGOING AND FURTHER TESTING IS PLANNED TO UNDERSTAND THE ROOT CAUSE OF THE ISSUE.