MalfunctionNBC

TRIAGE CARDIAC PANEL 25 TEST

Received May 22, 2015 · Event occurred Apr 28, 2015

Report 2027969-2015-00364 · MDR key 4792782

Device

Generic name

Cardiac Marker Test

Model number

97000HS

Lot number

W59043B

Product problems

  • Low Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

THE CUSTOMER CALLED AND ALLEGED A POTENTIAL LOW TRIAGE TROPONIN (TNI) RESULT IN COMPARISON TO THE VITROS LABORATORY ANALYZER RESULT. RESULTS ARE AS FOLLOWS: DATE: (B)(6) 2015, DEVICE: TRIAGE, TNI: <0.05. (B)(6) 2015, VITROS, 1.4. EDTA AND SERUM TUBE WAS DRAWN WITHIN 5 MINUTES OF EACH OTHER. THE VITROS CRITICAL VALUE >0.1. THERE WAS NO ADDITIONAL INFORMATION PROVIDED REGARDING THE PATIENT OR TESTING.

Additional Manufacturer Narrative

INVESTIGATION/CONCLUSION:THE CUSTOMER'S COMPLAINT OF DISCREPANT LOW TNI WAS NOT OBSERVED WITH IN-HOUSE TESTING ON RETAIN LOT W59043B. TESTING MET SPECIFICATIONS VALUES THAT ARE WITHIN THE PACKAGE INSERT CLAIM AND THE LOT PERFORMED AS EXPECTED.REVIEWED THE BATCH RECORD OF LOT W59043B FOUND NO NON-CONFORMANCES OR TESTING INVESTIGATION FORMS GENERATED DURING PRODUCTION OF THE LOT. THERE WAS NO PRODUCT DEFICIENCY ESTABLISHED AND NO CORRECTIVE ACTION IS REQUIRED.

Additional Manufacturer Narrative

(B)(4). INVESTIGATION PENDING.