inforMED
InjuryFKB

BUDGET MEDICAL - ICU MEDICAL

Received Jul 29, 2003 · Event occurred Jun 27, 2003

Report MW1029162 · MDR key 475473

Device

Generic name

T-connector Ped With Valve

Manufacturer

Budget Medical

Model number

*

Catalog number

*

Lot number

*

Product problems

  • Coagulation in Device or Device Ingredient
  • Use of Device Problem

Patient

12 MO

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

USED T-CONNECTOR FOR AN ARTERIAL LINE ON A PT. T-CONNECTOR WAS FOUND TO HAVE A CLOT FORMATION. FURTHER DISCUSSION WITH PHYSICIAN INDICATED THAT STAFF HAVE BEEN USING T-CONNECTOR IN THIS MANNER (AS AN ARTERIAL LINE CONNECTOR) WHEN NEITHER THE PACKAGING NOR THE MANUFACTURER HAD INDICATED NOT TO USE THE DEVICE IN THIS MANNER. PHYSICIAN IS AFRAID A LESS EXPERIENCED MD WILL NOT NOTICE THE CLOTTING AND FLUSH THE LINE INTO AN ARTERY, EMBOLIZING A CLOT, POTENTIALLY CAUSING STROKE AND FOR DEATH.