MalfunctionHAW
NDI PASSIVE SPHERES
Received Apr 2, 2015 · Event occurred Apr 9, 2013
Report 3002743211-2015-00018 · MDR key 4652026
Device
Generic name
Reflective Passive Marker Spheres
Manufacturer
Northern Digital Inc.Model number
8801075Lot number
1304291 / 1304261
Product problems
- Device Operates Differently Than Expected
- Output Problem
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
THIS EVENT WAS COMMUNICATED BY MEDTRONIC NAVIGATION WHO REPORTED THAT A SITE ((B)(6) HOSPITAL) HAD INSTRUMENTS THAT WOULD NOT TRACK. SITE REPLACED SPHERES ON INSTRUMENT AND THEN THE INSTRUMENT WAS ABLE TO TRACK. SITE REPORTED SPHERES LOOKED LIKE THEY WERE SCUFFED. SITE DISPOSED OF SPHERES AND DID NOT TAKE ANY PICTURES OF THE SPHERES. NO SERIOUS IMPLICATIONS TO THIS ISSUE WERE MENTIONED BY THE COMPLAINANT. NO PATIENT/USER INJURY OR OTHER SERIOUS HARM OCCURRED.