Device

Generic name

Saline Filled Testicular Prothesis

Manufacturer

Coloplast A/s

Model number

5206302400

Catalog number

450-1327

Lot number

4254879

Product problems

  • Deflation Problem

Patient

Not reported

  • No Code Available

Narrative

Description of Event or Problem

AS REPORTED TO COLOPLAST. THOUGH NOT VERIFIED, THE PATIENT WAS IMPLANTED WITH A TOROSA SALINE FILLED TESTICULAR PROSTHESIS. LATER THE PATIENT EXPERIENCED THAT THE PROTHESIS HAD DEFLATED. AN EXPLANT OF THE DEVICE WAS PERFORMED.

Additional Manufacturer Narrative

BECAUSE THE AVAILABLE INFORMATION PROVIDED INDICATED "LEAKING" MAY HAVE CONTRIBUTED TO THE DEVICE NOT WORKING AS INTENDED/DEVICE DEFLATION, AND BECAUSE NO SECONDARY INJECTION SITE WAS NOTED IN THE INJECTION PORT AND NO CRYSTALIZED RESIDUE WAS PRESENT IN THE DEVICE, QA DETERMINED THAT THE DEVICE WAS MOST LIKELY IMPLANTED WITHOUT BEING PROPERLY FILLED WITH SALINE ACCORDING TO THE IFU INSTRUCTIONS.