inforMED
MalfunctionDPW

MAQUET CARDIOPULMONARY AG

Received Mar 18, 2015

Report 8010762-2015-00255 · MDR key 4620299

Device

Generic name

Hl 20 Console Base

Model number

NI

Catalog number

70102.7652

Lot number

NA

Product problems

  • Noise, Audible

Patient

NI

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED, WHILE PERFORMING A FUNCTION CHECK A CLICKING SOUND OCCURED THAT WAS NOT PRESENT INITIALLY. AFTER OPERATING UNIT OVERNIGHT AT FULL 250 RPM'S CLICKING CEASED. UPON DISASSEMBLY I NOTICED THAT (UPPER) BELT WAS TWISTED AND THE (LOWER) BELT WAS BROKEN. (B)(4).

Additional Manufacturer Narrative

MAQUET MEDICAL SYSTEMS, USA SUBMITS THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE, MAQUET CARDIOPULMONARY (B)(4). PUMP PRODUCT NUMBER: MCP00703309. PUMP SERIAL NUMBER: (B)(4). THE ACTUAL DEVICE HAS NOT BEEN EVALUATED YET. A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED UPON RECEIPT OF ADDITIONAL INFO. ITEMS MARKED NI ARE UNKNOWN TO US AT THIS TIME.