inforMED
MalfunctionIPF

ZYNEX

Received Feb 11, 2015 · Event occurred Oct 3, 2013

Report 1723686-2015-00001 · MDR key 4574819

Device

Generic name

Nexwave

Manufacturer

Zynex Medical Inc

Product problems

  • Charred
  • Temperature Problem

Patient

Not reported

  • Electric Shock

Narrative

Description of Event or Problem

UNIT GETS HOT. USE TO BE ABLE TO USE ONE BATTERY EVERY 24HRS, NOW IT'S 3 BATTERIES EVERY 24 HRS. UNIT GETS SO HOT IT'S UNBEARABLE. PATIENT IS EXPERIENCING MORE SURGE SHOCKS NOW.

Additional Manufacturer Narrative

REPORTING DUE TO AN FDA FINDINGS DURING AUDIT IN 1/2015. INVESTIGATION AND RESULTS; UNIT TURNED ON, PASSED FINAL TESTING WITHOUT ANY PROBLEMS. LEFT UNIT RUNNING FOR 30 MINUTES AT 31MA ON BOTH CHANNELS (TENS MODE) AND HAD NO HEAT ISSUE. COULD NOT DUPLICATE PATIENTS COMPLAINT.

Remedial action

  • Replace