MalfunctionIPF

ZYNEX

Received Feb 11, 2015 · Event occurred Jun 26, 2014

Report 1723686-2015-00006 · MDR key 4574816

Device

Generic name

Truwave Plus

Manufacturer

Zynex Medical Inc

Product problems

  • Charred
  • Temperature Problem

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

UNIT IS BURNING AND PATIENT IS GETTING CHANNEL ELE ERRORS, HE USES THE TENS UNIT ON 25-30MA'S.

Additional Manufacturer Narrative

REPORTING DUE TO FDA FINDINGS DURING AUDIT IN 1/2015. INVESTIGATION AND RESULTS: UNIT ARRIVED IN IFC-LOWHIGH, 40 MINUTES. (B)(4). THIS UNIT DRAWS EXCESSIVE CURRENT; Q26 AND Q3 ARE DAMAGED DUE TO OVERHEATING. ELECTRODE WIRES AND POWER SUPPLY ARE FUNCTIONING PROPERLY.

Remedial action

  • Replace